Chapter 10
257
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1?
Description
Sets/Gets the Threshold1 value for Contact check. Threshold1 is a parameter which is
proportional contact resistance of Hp or Hc. It is recommended to use the default (initial)
value.
NOTE
The available range for Hp, Hc and Lc is 220
μ
F to 1 mF at 120 Hz and 22
μ
F to 100
μ
F at
1 kHz. The contact resistance for Lp cannot be detected by contact check feature for this
range.
Parameters
Query response
<numeric><newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 256
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 on page 258
Equivalent key
sequence
[
Meas Setup
] - CONT CHK- CC1 TH1
<numeric>
Description
Threshold1 value for Contact1.
Range
0 to 1
Initial value
0.1
Resolution
0.01