290
Chapter 10
Command Reference
E4981A SCPI Commands
:TEST:HANDler:STATus:OVLD
Syntax
:TEST:HANDler:STATus:OVLD {ON|OFF|1|0}
Description
This command sets the Handler Overload signal status for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (No Query)
NOTE
When
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Related commands
:TEST:HANDler:STATus:ALARm on page 286
:TEST:HANDler:STATus:INDex on page 288
:TEST:HANDler:STATus:NC on page 289
:TEST:HANDler:STATus:RDYTrig on page 291
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON
Sets Handler Overload signal status to ON.
OFF
Sets Handler Overload signal status to OFF.