Chapter 10
229
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe]:AVERage[:STATe]
Syntax
[:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description
Turns ON/OFF the averaging function.
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe]:AVERage:COUNt on page 228
Equivalent key
sequence
[
Meas Setup
] - AVG - ON/OFF
NOTE
Setting the averaging count with the front panel keys will automatically turn on the
function.
[:SENSe]:CORRection:CKIT:STAN1:FORMat
Syntax
[:SENSe]:CORRection:CKIT:STAN1:FORMat {GB|CPG}
[:SENSe]:CORRection:CKIT:STAN1:FORMat?
Description
Sets the parameter types of the OPEN correction data.
Parameters
Query response
{GB|CPG}<newline><^END>
Related commands
[:SENSe]:CORRection:DATA on page 236
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - OPEN - G-B/Cp-G
Description
ON or 1 (initial value)
Turns ON the averaging function.
OFF or 0
Turns OFF the averaging function.
Description
GB (initial value)
Specifies G as the primary parameter and B as the secondary
parameter.
CPG
Specifies Cp as the primary parameter and G as the secondary
parameter.