Chapter 10
187
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:CALCulate1:COMParator[:STATe]
Syntax
:CALCulate1:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator[:STATe]?
Description
Turns ON/OFF the comparator function.
This setup is interlocked with the ON/OFF state of the signal output of the handler
interface.
NOTE
Changing the measurement parameter will automatically turn off this function. To avoid
this, you have to write a program so that this command is executed after the execution of
the measurement parameter setup command (
Parameters
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
Equivalent key
sequence
[
Meas Setup
] - LIMIT TABLE - COMP
Description
ON or 1
Turns ON the comparator function.
OFF or 0 (initial value)
Turns OFF the comparator function.