Chapter 10
233
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe]:CORRection:COLLect[:ACQuire]
Syntax
[:SENSe]:CORRection:COLLect[:ACQuire] {STAN1|STAN2|STAN3}
Description
Measures the correction data for OPEN/SHORT/LOAD correction and turns on the
correction function.
For Open and Short correction, the correction data is measured for all measurement
frequencies. For load correction, the correction data is measured for the measurement
frequency used in executing the command (set with the
command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
command), this is measured as the correction
data for the selected channel when executing the command (selected with the
[:SENSe]:CORRection:MULTiple:CHANnel
command). (No query)
Parameters
Related commands
[:SENSe]:CORRection:OPEN[:STATe] on page 244
[:SENSe]:CORRection:SHORt[:STATe] on page 245
[:SENSe]:CORRection:LOAD[:STATe] on page 238
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - OPEN - MEAS OPEN
[
Meas Setup
] - CORRECTION - SHORT - MEAS SHORT
[
Meas Setup
] - CORRECTION - LOAD - MEAS LOAD
Description
STAN1
Specifies the OPEN correction.
STAN2
Specifies the SHORT correction.
STAN3
Specifies the LOAD correction.