Chapter 10
Command Reference
E4981A SCPI Commands
:DATA:REFerence1:DATA
Syntax
:DATA:REFerence1:DATA <numeric>
:DATA:REFerence1:DATA?
Description
Sets or reads out the reference value for the primary parameter used in the deviation
measurement mode.
NOTE
The functionality of this command is same as
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Related commands
:DATA:REFerence2:DATA on page 205
Equivalent key
sequence
[
Meas Setup
] - REF A or B
:DATA:REFerence1:FILL
Syntax
:DATA:REFerence1:FILL
Description
Execute a single measurement and set the measured value into the primary and secondary
reference values for deviation. This command is same as
. (No Query)
Related commands
:DATA:REFerence2:FILL on page 205
Equivalent key
sequence
[
Meas Setup
] - REF A - MEASURE
<numeric>
Range
-99.9999E9 to 99.9999E9
Initial value
0