![Keysight Technologies E4981A Programming Manual Download Page 51](http://html1.mh-extra.com/html/keysight-technologies/e4981a/e4981a_programming-manual_1961965051.webp)
Chapter 3
49
Setting Up Measurement Conditions and Display
Setting Up Beep
3.
Se
ttin
g
up
Me
asur
eme
nt
Co
ndition
and
Displa
y
Setting Up Beep
Setting up conditions to make a beep sound
To turn ON/OFF the beep sound, use one of the following commands. You can use either of
these two commands since both functions are identical.
•
:CALCulate1:COMParator:BEEPer[:STATe] on page 176
•
:SYSTem:BEEPer:STATe on page 269
To generate the beep sound regardless of ON/OFF setting, use
:SYSTem:BEEPer[:IMMediate] on page 268
. To change the tone of beep sound from
five different types of available tones, use
:SYSTem:BEEPer:TONE on page 269
.
To set the conditions to make a beep sound according to the comparator sorting result, use
the following command.
•
:CALCulate1:COMParator:BEEPer:CONDition on page 175
The table below shows the relationship between the settings made by the above commands
and the condition to make a beep.
Making a beep sound
To make a beep sound, use the following command:
•
:SYSTem:BEEPer[:IMMediate] on page 268
Setting of
or
Setting of
Condition to make a beep
OFF
———
Never make a beep sound
ON
FAIL
•When wrong key
operation is
performed.
•When an error,
alarm, or other
message is output.
•When the sorting
judgment result of the
comparator is
OUT_OF_BIN, OVLD,
LOWC_OR_NC or
AUX_BIN.
PASS
•When the sorting
judgment result of the
comparator is within
BIN1 to BIN9.