Chapter 10
205
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:DATA:REFerence2:DATA
Syntax
:DATA:REFerence2:DATA <numeric>
:DATA:REFerence2:DATA?
Description
Sets or reads out the reference value for the secondary parameter used in the deviation
measurement mode.
NOTE
The functionality of this command is same as
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Related commands
:DATA:REFerence1:DATA on page 204
Equivalent key
sequence
[
Meas Setup
] - REF A or B
:DATA:REFerence2:FILL
Syntax
:DATA:REFerence2:FILL
Description
Execute a single measurement and set two measured value into each of the reference values
for deviation. This command is same as
:DATA:REFerence1:FILL on page 204
Query)
Related commands
:DATA:REFerence1:FILL on page 204
Equivalent key
sequence
[
Meas Setup
] - REF B- MEASURE
<numeric>
Range
-99.9999G to 99.9999G
Initial value
0