![Keysight Technologies E4981A Programming Manual Download Page 267](http://html1.mh-extra.com/html/keysight-technologies/e4981a/e4981a_programming-manual_1961965267.webp)
Chapter 10
265
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:SOURce:VOLTage:MODE
Syntax
:SOURce:VOLTage:MODE {CONTinuous|SYNChronous}
:SOURce:VOLTage:MODE?
Description
Determines whether to use the synchronous source function (to output the measurement
signal only during measurement) or not (to always output the measurement signal). The
synchronous source function lets you set the source delay time with the
“:TRIGger[:SEQ1]:DELay” on page 297
, to suspend the signal output during the waiting
time after a trigger is generated.
Parameters
Query Response
{CONT|SYNC}<newline><^END>
Equivalent key
sequence
[
Meas Setup
] - SYNC SRC- ON|OFF
:STATus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
Reads out the value of the Operation Status Condition register. (Query only)
Query response
{numeric}<newline><^END>
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
CONTinuous
Always outputs the measurement signal.
SYNChronous