258
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2?
Description
Sets/Gets the Threshold2 value for Contact check. Threshold2 is a parameter which is
proportional contact resistance of Lp or Lc. The recommended setting value is described in
the
Operation Manual
.
NOTE
The
contact failure at Lp can not be detected at the range of 220 uF to 1 mF at 120 Hz and
22 uF to 100 uF at 1 kHz. At the same range, the threshold1 is used instead of this for Lc.
(Therefore, the threshold2 is not used at the range).
Parameters
Query response
<numeric><newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 256
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 on page 257
Equivalent key
sequence
[
Meas Setup
] - CONT CHK- CC1 TH2
<numeric>
Description
Threshold2 value for contact 1
Range
0 to 1
Initial value
1
Resolution
0.01