:CALCulate1:COMParator:MODE
Syntax
:CALCulate1:COMParator:MODE {ABS|DEV|PCNT}
:CALCulate1:COMParator:MODE?
Description
Determines how to specify the limit range of the primary parameter for the comparator
function.
Parameters
Query response
{ABS|DEV|PCNT}<newline><^END>
Related commands
:CALCulate1:COMParator:PRIMary:NOMinal on page 184
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT] on page 182
Equivalent key
sequence
[
Meas Setup
] - LIMIT TABLE - Delta Mode
Description
ABS
(initial value)
Specifies the limit border value in an absolute value (absolute mode).
DEV
Specifies the border value relative to the reference value
*1
as an
absolute value (absolute tolerance mode).
PCNT
Specifies the border value relative to the reference value as a
percentage of the reference value
(percent tolerance mode).
*1.Use the
:CALCulate1:COMParator:PRIMary:NOMinal
command to set the reference
value.