Chapter 10
215
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD:DATA
Syntax
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD:DATA <numeric>
:DISPlay[:WINDow]:TEXT2[:DATA]:FMSD:DATA?
Description
Sets the value of the highest digit of the secondary parameters in the fixed point display.
The parameter is always fixed to the same value, regardless of this setting, if the following
conditions are met:
• Deviation is displayed as a percentage in the deviation measurement mode.
• The resulting D, Q and % of the secondary parameter is displayed.
Parameters
If one of the settable values is not specified for the parameter, the minimum possible value,
which is larger than the specified parameter value, is set. If the specified parameter exceeds
the maximum value, the maximum value is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD[:STATe] on page 214
:CALCulate1:MATH:EXPRession:NAME on page 189
:CALCulate1:MATH:STATe on page 190
:CALCulate2:MATH:EXPRession:NAME on page 192
:CALCulate2:MATH:STATe on page 193
Equivalent key
sequence
[
Display Format
] - <Secondary Measurement Parameter i.e D/Q/G/Rp/Rs> - D.P. POS
INCR+|D.P. POS DECL-
<numeric>
Description
The value of the highest digit of the secondary parameter.
Range
1a|10a|100a|1f|10f|100f|1p|10p|100p|1n|10n|100n|1u|10u|100u|1
m|10m|100m|1|10|100|1k|10k|100k|1M|10M|100M|1G|10G|100
G|1T|10T|100T|1P|10P|100P|1E
Initial value
1n