Chapter 10
193
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:CALCulate2:MATH:STATe
Syntax
:CALCulate2:MATH:STATe {ON|OFF|1|0}
:CALCulate2:MATH:STATe?
Description
Determines whether to use the function (deviation measurement mode) that displays the
secondary parameter measurement result as deviation from the reference value (set using
the
NOTE
Changing any of the measurement parameters will automatically disable this function. To
avoid this, you have to write a program so that this command is executed after the
execution of the measurement parameter setup command (
Parameters
Query response
{1|0}<newline><^END>
Related commands
:CALCulate2:MATH:EXPRession:NAME on page 192
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
:CALCulate1:MATH:STATe on page 190
Equivalent key
sequence
[
Meas Setup
] - DEV B - OFF
:CALCulate3:MATH:STATe
Syntax
:CALCulate3:MATH:STATe {ON|OFF|1|0}
:CALCulate3:MATH:STATe?
Description
This is a dummy command for 4268A/4288A compatibility. The current monitor function
is always set to ON in E4981A. The query response is always 1.
Query response
{1}<newline><^END>
Related commands
:CALCulate4:MATH:STATe on page 194
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1
Enables the deviation measurement mode.
OFF or 0 (initial value)
Disables the deviation measurement mode (i.e., displays
the measurement result as an absolute value).