192
Chapter 10
Command Reference
E4981A SCPI Commands
:CALCulate2:MATH:EXPRession:NAME
Syntax
:CALCulate2:MATH:EXPRession:NAME {DEV|PCNT}
:CALCulate2:MATH:EXPRession:NAME?
Description
Specifies the expression of the secondary parameter used when displaying the
measurement result in the deviation measurement mode.
Parameters
Query response
{DEV|PCNT}<newline><^END>
Related commands
:CALCulate2:MATH:STATe on page 193
:CALCulate1:MATH:EXPRession:NAME on page 189
:DATA[:DATA] on page 206
Equivalent key
sequence
[
Meas Setup
] - DEV B - delta ABS|delta %
Description
DEV
(initial value)
Displays the result as the difference between the measurement
value and the reference value
*1
(measurement value - reference
value).
PCNT
Displays the difference between the measurement value and the
reference value
*2
of the reference value.
*1.Use the
command to set the reference value.
*2.(Measurement value - reference value) / reference value
∞
100