208
Chapter 10
Command Reference
E4981A SCPI Commands
When specifying BUF3 as parameter (Comparator: ON)
{numeric 1},{numeric 2},{numeric 3},{numeric 4},
…
,{numeric (N
∞
4)}<newline><^END>
The N sets of measurement data (a data set consists of the measurement status, measured
value, and comparator sorting result) fed into the data buffer are read out in the order of
measurement.
Where N is the number of measurement points specified with the
command and n is an integer between 1 and N:
Related commands
:DATA:FEED:BUF1 on page 194
:DATA:FEED:CONTrol[:STATe] on page 198
:DATA:POINts:BUF1 on page 200
:DATA:POINts:BUF2 on page 201
:DATA:POINts:BUF3 on page 202
:CALCulate3:MATH:STATe on page 193
:CALCulate4:MATH:STATe on page 194
:FETCh? on page 217
:FORMat[:DATA] on page 222
Equivalent key
sequence
No equivalent key is available on the front panel.
{numeric (n
−
1)
∞
3+1} The measurement status in the n-th measurement (an integer
between 0 and 2).
{numeric (n
−
1)
∞
3+2} The measured value of the primary parameter in the n-th
measurement. If overload is detected, 9.9E37 is outputted.
{numeric (n
−
1)
∞
3+3} The measured value of the secondary parameter in the n-th
measurement. If overload is detected, 9.9E37 is outputted.
{numeric (n
−
1)
∞
3+4}
The comparator sorting result in the n-th measurement (an
integer between 0 and 11). The output (output value is 11) is
produced even if the comparator is off.