Chapter 10
201
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:DATA:POINts:BUF2
Syntax
:DATA:POINts:BUF2 <numeric>
:DATA:POINts:BUF2?
Description
Specifies the size of data buffer 2 in number of measurements. Executing this command
returns the pointer to the specified data buffer (the location to feed measurement data) back
to the start.
NOTE
The functionality of this command is same as
with value of BUF2.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric} <newline><^END>
Related commands
:DATA:FEED:BUF2 on page 195
:DATA:FEED:CONTrol[:STATe] on page 198
Equivalent key
sequence
No equivalent key is available on the front panel.
<numeric>
Description
Number of measurements for the data buffer specified with BUF2
Range
1 to 200
Initial value
200
Resolution
1