Chapter 10
285
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:TEST:HANDler:MODE
Syntax
:TEST:HANDler:MODE {ON|OFF|1|0}
:TEST:HANDler:MODE?
Description
This command sets the Handler Comparator Function value for test purpose and can be
used to check/troubleshoot the Test pin signal of Handler interface.
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Query response
{1|0}<newline><^END>
Related commands
:TEST:HANDler:BIN on page 283
:TEST:HANDler:COMP on page 284
:TEST:HANDler:KEYLock? on page 284
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1
Sets Handler mode to ON.
OFF or 0 (initial value)
Sets Handler mode to OFF.