292
Chapter 10
Command Reference
E4981A SCPI Commands
:TEST:REAR:TRIGger?
Syntax
:TEST:REAR:TRIGger?
Description
This command gets the BNC Trigger signal for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (Query Only)
NOTE
When
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Query response
{1|0}<newline><^END>
Related commands
:TEST:HANDler:TRIGger? on page 291
Equivalent key
sequence
No equivalent key is available on the front panel.
:TEST:SCANner:CH?
Syntax
:TEST:SCANner:CH?
Description
This command gets the Scanner channel number for test purpose and can be used to
check/troubleshoot the Test pin signal of Scanner interface. (Query Only)
NOTE
When
is set to ON, the interface pin signal value can be
controlled/read by :TEST:SCAN:xxxx commands. After completing the test,
should set to OFF so that scanner interface pin signal is generated
according to the actual measurement result.
Query response
<numeric><newline><^END>
Related commands
:TEST:SCANner:EOM on page 293
:TEST:SCANner:INDex on page 294
:TEST:SCANner:MODE on page 295
:TEST:SCANner:TRIGger? on page 296
:TEST:SCANner:VALid? on page 296
Equivalent key
sequence
No equivalent key is available on the front panel.