3.5 Input Channel Setting
75
3
Chapte
r 3 Me
asu
remen
t Proced
ure
4. Vertical axis
(Voltage axis)
Zoom
Vertical axis (voltage axis) zoom-up or zoom-down settings can be made sepa-
rately for each channel. The settings will be used for display.
Zooming is carried out using the zero position as reference. The measurement
resolution does not change.
See:
"6.5.3 Magnifying and Compressing Vertical Axis (Voltage Axis)" (p.129)
To achieve a user-specified zoom setting, the Variable function is used. By
reversing plus/minus, the waveform can be inverted.
See:
"7.5 Variable Function (Setting the Waveform Display Freely)" (p.153)
"7.7 Inverting the Waveform (Invert Function)" (p.157)
5. Vernier
Fine adjustment of input voltage can be performed arbitrarily on the Waveform
screen (display only). When recording physical values such as noise, temperature
and acceleration using sensors, amplitude can be adjusted to facilitate calibration.
See:
"7.6 Fine Adjustment of Input Values (Vernier Function)" (p.156)
6. Zero Position
Sets the 0 V level display position. If the 0 V input level has shifted, perform
zero-adjust.
See:
"2.6 Adjusting the Zero Position (Zero-Adjust)" (p.57)
See:
"2.7 Performing Calibration (When Mounting MR8990)" (p.58)
If the zero position of the 8969 and U8969 Strain Unit is out of alignment, perform
auto balance.
See:
"7.9.4 Settings for the 8969 and U8969 Strain Unit" (p.163)
The zero position is as shown in the illustration below.
(Example: 8966 Analog Unit)
• Simply moving the display position will not apply an offset to the input.
• Zoom of the vertical axis direction (voltage axis) is based on the zero position.
• The voltage range displayed in the waveform screen changes with zero posi-
tion and zoom of the vertical axis (voltage axis) but the measurable range does
not change.
A/D Data
2047
0
-2047
A/D Data
0 V
2000 LSB
0 %
50 %
100 %
Display screen
(Zero position: 0%)
Display screen
(Zero position: 50%)
Display screen
(Zero position:100%)
<Zoom factor ×1>
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......