7.9 Making Detailed Settings for Input Modules
167
6
Chapte
r 7 Utility Functions
7
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (p.159)
Mode
There is no need to change the setting since it is set when the clamp sensor is
automatically recognized.
Select
*; When Model CT6846 or CT6865 is connected to Model 8971 Current Unit through Model 9318
Conversion Cable, the instrument recognizes that as an 500 A AC/DC sensor is connected;
set the scaling ratio to 2.00.
Range
Select
7.9.6 Settings for the 8971 Current Unit
Selections
Description
20A/2V
Sets this option when Model 9272-10 (20 A range), 9277, or
CT6841 Clamp Sensor is connected. (default setting)
200A/2V
Sets this option when Model 9272-10 (200 A range), 9278,
CT6863, or CT6843 Clamp Sensor is connected.
50A/2V
Sets this option when Model CT6862 Clamp Sensor is connected.
500A/2V
Sets this option when Model 9279, 9209, CT6544, CT6845,
CT6846*, or CT6865* Clamp Sensor is connected.
Selections
Description
DC
Current measurement (Default setting)
RMS
RMS current measurement
• 8971 Current Unit cannot be used with MR8741.
• With MR8740, up to four 8971 Current Units can be used.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......