12.3 Setting FFT Analysis Conditions
252
The averaging function calculates the average of the values obtained from multiple measurements of a peri-
odic waveform. This can reduce noise and other non-periodic signal components. Averaging can be applied
to a time-domain waveform or to a spectrum.
12.3.7 Averaging Waveforms
1
Enable averaging.
Move the flashing cursor to the
[Average]
item.
Select
2
Select the count for averaging.
Move the flashing cursor to the
[Number]
item.
Select the number of measurements to be averaged.
Setting range: 2 to 10,000
Off
Averaging is disabled. (default setting)
Linear
(Time axis)
Time-domain waveforms are summed and averaged
and then calculation is performed.
Expo.
(Time axis)
The exponential mean of time-domain waveforms is
determined and then calculation is performed.
Linear
(Frequency)
Frequency-domain waveforms are summed and aver-
aged and then the calculation result is output.
Expo.
(Frequency)
The exponential mean of frequency-domain wave-
forms is determined and then the calculation result is
output.
Peak
(Frequency)
The maximum value of frequency-domain waveforms
is retained.
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Status]
sheet
About averaging calculation formulas
See: "Averaging" (p.A19)
When averaging and auto saving are
enabled at the same time
Data is saved after the specified count of
values have been averaged.
See:
"Trigger Modes and Averaging"
(p.254)
1
2
• After measuring with averaging enabled, display is not available when the
channel is changed. Also, when the analysis mode is changed, the analysis
modes that can be displayed are limited.
• When averaging is performed with the analysis mode disabled (Off), no trace
is displayed when the analysis mode is changed after measurement.
•
When
[Format]
is set to
[Running spectrum]
,
[Average]
cannot be set.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......