![Hioki MR8740 Instruction Manual Download Page 190](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324190.webp)
7.9 Making Detailed Settings for Input Modules
178
Method
Selects the control method for waveform output.
Select
Settings when selecting arbitrary waveform
Data name
Register/delete the data of the outputted waveform. Maximum eight waveforms
can be registered.
Filter
Filter the output waveform.
Clock Freq
Set the clock frequency of the D/A converter that generates waveforms.
The period and frequency of the output waveform are displayed on the lower
level.
Delay
Set the delay for waveform generation. When the delay value is positive, the
phase is progressive.
Loop
Set the repetition count for generating waveforms.
Amp Adj
Set the amplitude level for waveform output.
Offset
Set the offset voltage for waveform.
Selections
Description
Manual
Restricts control of signal output to Signal Generation screen.
Sync.
Augments manual control with signal output in synchronization with
the start and end of measurement.
START
key: Starts output when measurement starts.
STOP
key: Stops output when measurement stops.
Keys
Augments manual control by allowing signal output to be manipu-
lated using the instrument’s keys.
START
key: Starts output.
STOP
key: Stops output.
Manual Trigger key: Pauses output.
Off (Default setting) to 1 MHz
-250,000 data to 250,000 data (Default settings: 0)
1 to 50,000,
∞
(Default setting:
∞
)
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......