2.1 Install an input module
36
Input modules specified at the time the instrument is ordered are supplied preinstalled. Use the fol-
lowing procedures to add or replace input modules, or to remove them from the instrument.
2.1 Install an input module
Preparations
• To avoid electric shock accident, before removing or replacing an input mod-
ule, confirm that the instrument is turned off and that the connection cords are
disconnected.
• The mounting screws must be firmly tightened or the input module may not
perform to specifications, or may even fail.
When an input module is not used
• To avoid the danger of electric shock, never operate the instrument with an
input module removed. To use the instrument after removing an input module,
install a blank panel over the opening of the removed module.
• To avoid damaging input modules, do not touch the input module connectors
on this instrument.
• Measurements made without a blank panel installed may fail to meet specifica-
tions because of temperature instability within the instrument.
For information on analog channel accuracy when using logic channels, see "7.9
Making Detailed Settings for Input Modules" (p.159).
Required item: One Phillips-head screwdriver
1
Turn the instrument's
POWER
switch Off.
2
With attention to the orientation of the input module,
insert it firmly all the way in.
Make certain that the labels on the input module's panel
face the same direction as the labels on the right side of
the instrument.
3
Using the Phillips screwdriver, tighten the two input
module mounting screws.
Installing an input module
Front Side
Handle
(Example:8966)
Handle
Required item: One Phillips-head screwdriver
1
Turn the instrument’s
POWER
switch Off.
2
Remove any cables or thermocouples connected to
the input module.
3
Remove the power cord.
4
Using the Phillips screwdriver, loosen the two input
module mounting screws.
5
Grasp the handle and pull the module out.
Removing an input module
Front Side
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......