11.1 Recording Settings
239
10
Ch
apter 1
1
Memory Division
Function
11
11.1 Recording Settings
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Memory Div]
sheet
1
Enable the Memory Division function.
Move the flashing cursor to the
[Memory Div]
item.
Select
[On]
.
2
Set the number of divisions.
Move the flashing cursor to the
[Division]
item.
Set the number of blocks for division.
Default setting: 4
3
Set the recording length.
(This is linked to the recording length setting on the
[Status]
sheet.)
Move the flashing cursor to the
[Shot]
item.
Set the recording length.
The maximum recording length and number of divisions are determined
automatically according to memory capacity and the number of chan-
nels used.
Setting range
:
"Appendix 2.4 Maximum record length and number of di-
visions (Memory division function)"(p.A7)
4
Set the start block.
Move the flashing cursor to the
[Start Block]
item.
Set the block number at which to start recording.
Default setting: 1
5
Set the Used Block number.
Move the flashing cursor to the
[Use Blocks]
item.
Set the number of blocks to use.
Default setting: 1
Off
Memory Division is disabled.(default setting)
On
Memory Division is enabled.
About Recording
When a fast timebase is selected, display-
ing and saving operation are not available
while measuring.
Selecting the display screen for auto sav-
ing lengthens dead time.
Use Blocks (Purple)
Memory Division, Waveform Calcula-
tion, and Roll Mode cannot be enabled
at the same time.
When the number of divisions is 32, the
Start Block is 5 and the Used Block num-
ber (number of blocks to use) is 20
No. of Divisions
Start Block
3
1
2
4
5
To display any block on the waveform screen when finished measuring:
Set the number of blocks to display (p.240).
(This can also be set on the Waveform screen (p.134).)
To display overlaid waveforms:
Set the number of blocks for reference (p.240).
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......