7.9 Making Detailed Settings for Input Modules
168
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (p.159)
Mode
Switches between voltage measurement and RMS measurement.
Select
Response
Response can be set to three speeds: Fast, Normal and Slow.
Normally set to
[Fast]
, this can be changed to
[Normal]
or
[Slow]
to stabilize the
display when measuring low frequencies, or when severe fluctuations are
present.
Select
7.9.7 Settings for the 8972 DC/RMS Unit
Selections
Description
DC
Voltage measurement (default setting)
RMS
RMS measurement
Selections
Description
Fast
Sets the response time to about 100 ms. (default setting)
Normal
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 s.