![Hioki MR8740 Instruction Manual Download Page 381](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324381.webp)
18.3 Error Messages
369
C
hap
ter 18
M
aintena
nce and
Service
14
18
A screen message appears whenever an error occurs. In either case, take the remedial action indicated.
A beep may sound if the beeper setting on the
[Environment]
sheet is
[Warn]
or
[Warn+Action]
.
See:
"Chapter 14 System Environment Settings" (p.305)
18.3 Error Messages
Message
Displayed Message
Number (Msg No.)
Warning Display
Appears just once when an error
occurs. Disappears within a few
seconds.
Also disappears when clicking
the mouse.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
10
Mount media.
Set the recording media.
"2.3 Recording Media Prep-
aration" (p.52)
11
Illegal format.
Wrong recording format. Reformat the media.
"2.3.2 Formatting Storage
Media" (p.53)
12
Can not write.
Write-protection is enabled on the storage
media. Disable it.
-
13
Disk full while accessing the file.
Saving is not possible because of insufficient
space on the storage media. Delete files or re-
place the storage media. If measuring,
Stop measurement, then replace the storage
media.
"4.4.4 Deleting Files & Fold-
ers" (p.104)
14
File is read only.
File is read only. Deletion is not permitted.
-
15
Access to the file was denied.
An internal fault may have occurred in the in-
strument. Turn the instrument off and back on
-
16
The file name already exists: cannot
save.
Change the file name.
"4.4.6 Renaming Files &
Folders" (p.105)
17
The folder name already exists.
Change the folder name.
18
Not enough free space in the folder to
create file.
Either delete files in the saving destination
folder, or change to another saving destina-
tion.
-
19
Folder not empty.
The folder is not empty, be sure you want to
delete.
-
20
The maximum length of a file name, in-
cluding its path, is 255 characters.
File names including the path must not exceed
255 characters.
-
21
Internal error.
An internal error occurred. Check the storage
media.
-
22
No waveform data to save.
Acquire waveform data.
-
24
There is no calculation result.
There is no calculation result. Save results af-
ter performing calculation.
"Chapter 9 Numerical Cal-
culation Functions" (p.207)
26
Invalid folder path.
Route cannot be specified.
-
27
When executing, select
[No]
as the save
property.
As other dialogs are displayed, selective sav-
ing cannot be carried out. Either set the selec-
tive saving during execution to
[No]
or close
the dialog and re-save.
"4.2.3 Saving Data Selec-
tively (SAVE)" (p.91)
28
Pressed key invalid (waveform evalua-
tion).
Unable to perform operation because wave-
form evaluation is enabled. Turn off waveform
evaluation.
"13.3 Setting the Waveform
Evaluation Mode" (p.296)
29
The record length was restricted.
-
-
31
A/B cursor positions invalid.
The A/B cursors overlap. Check the cursor po-
sitions.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......