![Hioki MR8740 Instruction Manual Download Page 210](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324210.webp)
8.6 Synchronizing Between Blocks (MR8740 Only)
198
The input signal from another block can be used as a trigger source.
Use this to synchronize between block I and block II of MR8740 Memory HiCorder.
8.6 Synchronizing Between Blocks
(MR8740 Only)
Procedure
To open the screen: Right-click and select
[DISP]
→
Waveform screen
→
Right-click and select
[TRIG.SET]
→
Trigger settings window
1
Enable synchronization between blocks.
Move the flashing cursor to the
[Sync blocks]
item.
Select
Off
Synchronization between blocks is not performed.
(default setting)
On
Synchronization between blocks is performed.
• To synchronize block I and block II, trigger synchronization needs to be set for
both blocks.
• The sampling offset between blocks is within 3 samples or 1 us.
If the sampling differs between both blocks, the offset will be for the slow sam-
pling side.
• Synchronizing between blocks of the U8793, MR8790, or MR8791 cannot be
enabled.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......