![Hioki MR8740 Instruction Manual Download Page 333](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324333.webp)
15.2 Performing Remote Operations on the Instrument (Use an Internet Browser)
321
12
Ch
apter 15
Connection
to a Compute
r
14
15
Acquiring data by communication commands ______________________
Transferring storage data (measurement values)
Use
:MEMory:VDTA? A
.
Transferring storage data in binary format
Use
:MEMory:BDATA? A
.
Procedure
1
:MEMory:MAXPoint?
The number of data samples stored in the memory of the instrument is acquired.
Check that storage data exists.
2.
:MEMory:POINt CH$,A
The output point of the storage data is specified.
(If you want to acquire the data from the beginning, change the value of A to 0.)
3.
:MEMory:VDATA? A
Just the number of data samples of the storage data (measurement values) specified with A is transferred
from the instrument to the computer. The output point is also incremented by the number of data samples.
• From 1 to 80 values can be acquired at one time with the
VDATA
command. If
you want to acquire more values than that, send the command in step
3
repeatedly.
• The maximum number of data samples that can be acquired is the value
acquired with the
MAXPoint
command.
Procedure
1
:MEMory:MAXPoint?
The number of data samples stored in the memory of the instrument is acquired.
Check that storage data exists.
2.
:MEMory:POINt CH$,A
The output point of the storage data is specified.
(If you want to acquire the data from the beginning, change the value of A to 0.)
3.
:MEMory: BDATA? A
Just the number of data samples of the storage data (binary) specified with A is transferred from the instru-
ment to the computer. The output point is also incremented by the number of data samples.
• From 1 to 400 values can be acquired at one time with the
BDATA
command.If
you want to acquire more values than that, send the command in step
3
repeatedly.
• The maximum number of data samples that can be acquired is the value
acquired with the
MAXPoint
command.
• #0 (indicates binary format) is added to the beginning of the output data.
• After #0, the storage data is transmitted two bytes at a time.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......