7.9 Making Detailed Settings for Input Modules
170
Select
Selections
Description
Off
Calibration and synchronization are not performed.
On
Performs calibration and synchronization.
Synchroniza-
tion
Performs only synchronization between channels.
• The calibration time is approximately 150 ms. That period becomes a waiting
time in which measurement is not performed.
• If synchronization between channels is performed, a signal to stop integration
is sent to each unit when measurement is started and the wait process is per-
formed until the first integration finishes. The wait time required for this process
is (10 ms + integration time
*
).
*The integration time varies depending on the NPLC setting.
If synchronization is not performed, the above wait time will also be required
for measurement performed immediately after the settings of the Digital Volt-
meter Unit are changed. However, there will be no wait time when measure-
ment is performed with the same settings.
• When this is set to OFF (default setting), perform calibration manually.
See:
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......