![Hioki MR8740 Instruction Manual Download Page 268](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324268.webp)
12.3 Setting FFT Analysis Conditions
256
Select the type of FFT analysis, channel(s), waveform display color and x and y axes.
12.3.9 Analysis Mode Settings
1
Select the FFT analysis mode.
Move the flashing cursor to the
[Analyze]
column of the Analysis No.
to se
t.
Select
(When [List] is selected, a list of calculation types appears.)
2
Select whether to display the waveform, and its color.
Move the flashing cursor to the
[Wave]
column.
Select whether the waveform is to be displayed (On) or not, and its
color if displayed.
Analysis No.
Settings can be made from the
dialogs, or copied from another
Analysis No. (p.158)
Analysis
Type
1
2
Analysis Setting
Contents
Channel for
Analysis
3
4
5
Display
Color
X/Y Axes
Display
Parameter
Off
No analysis.
(default setting)
Storage Waveform
Histgram
Linear Spectrum
RMS Spectrum
Powre Spectrum
Powre Spectrum
density*
LPC (density)*
Transfer Function
Cross Power Spectrum
Impulse Response
Coherance Function
Phase Spectrum
Auto-Correlation
Cross-correlation
1/1 Octave*
1/3 Octave*
*Not available with external sampling enabled.
See:
"12.8.2 Analysis Mode Functions" (p.289)
Procedure
To open the screen: Right-click and select
[STATUS]
→
[Status]
sheet
See:
To set from the Waveform screen (p.261)
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......