6.6 Monitoring Input Levels (Level Monitor)
131
5
Chap
ter 6 W
aveform Scree
n Mo
nitoring an
d Analysis
6
Input values can be monitored as numerical values in the same way as with a DMM (digital multim-
eter).
6.6.2 Numerical Value Monitoring (DMM Display)
Procedure
To open the screen: Right-click and select
[DISP]
→
Left-click the
[DMM]
icon
Left-click the
[DMM]
icon.
Left-click the
[DMM]
icon on the top right of the Waveform
screen.
The display changes to numerical value (DMM) display.
To return to the Waveform screen
Left-click
[Back to waveform]
in the display menu on the right of the
screen.
To pause (hold) display
Left-click [
HOLD ON/OFF]
in the display menu on the right of the screen.
The HOLD mark appears at the top of the screen to indicate that the
screen is in the HOLD state.
Click [
HOLD ON/OFF]
in the HOLD state to cancel hold.
Zero adjust
Zero adjustment (calibration for MR8990) can also be performed in the
numerical value (DMM) display screen. Left-click
[Exec Zero-Adjust]
in
the display menu on the right of the screen.
Display menu
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......