![Hioki MR8740 Instruction Manual Download Page 215](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324215.webp)
8.10 Setting Trigger Timing
203
7
Cha
pter 8 T
rigger Set
tings
8
This applies to the Memory function only.
Set waveform recording operation when a trigger event occurs.
Description
About trigger timing
The selected trigger mode determines how recording stops.
Example: When the trigger type is Level Trigger, Level = 0.000 V, and Slope =
↑
(rising)
8.10 Setting Trigger Timing
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings window
Move the flashing cursor to the
[Timing]
item.
Select
Start
Start recording when a trigger event occurs, and stop after
the specified recording length.(default setting)
Stop
Start recording when the
START
key is pressed, and stop
when a trigger event occurs.
Start & Stop
Record the interval from one trigger event until the next
trigger event.
[Start]
[Stop]
[Start & Stop]
T
Recording
START
key
T
Recording
T
T
Recording
Recording
Starts
Recording starts when a trigger
event occurs
Recording starts when you press
the
START
key
Recording starts when a Start trig-
ger event occurs
Recording
Stops
With
[Single]
trigger mode
Recording stops after data has
been acquired for the specified
recording length
Recording stops when a trigger
event occurs
Recording stops when a Stop trig-
ger event occurs
If no trigger event occurs before the specified recording length elapses:
[Stop]
or
[Start & Stop]
: Recording stops after data has been acquired for the specified recording length
With
[Repeat]
trigger mode
The Trigger Wait state begins af-
ter data has been acquired for the
specified recording length
When another trigger event oc-
curs, data is again acquired for the
specified recording length, then
Trigger Wait resumes (repeats)
When a trigger event occurs, re-
cording stops and then starts
again (repeats)
When a trigger event occurs, re-
cording stops and the Trigger Wait
state resumes. When another trig-
ger event occurs, recording con-
tinues until the next trigger occurs
(repeats)
If no trigger event occurs before the specified recording length elapses:
[Stop]
: After data is acquired for the specified recording length, recording restarts. This repeats until
a trigger event occurs.
[Start & Stop]
: The Trigger Wait state begins after data has been acquired for the specified recording length
(Start Trigger)
[Start]
[Stop]
[Start & Stop]
Records for specified record-
ing length
Press
START
key to record
Records until a trigger occurs
T
T
START
key
Recording starts when a Start
trigger event occurs Records
until a Stop trigger occurs
T
T
The above sequences repeat when the trigger mode is
[Repeat]
.
0 V
0 V
0 V
Setting of
[Timing]
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......