3.3 Pre-Measurement Inspection
62
The following steps should be performed before measurement.
3.3 Pre-Measurement Inspection
Before using the instrument the first time, verify that it operates normally to ensure that the no
damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki
representative.
Do not use if damage is present,
as you could receive an electric
shock. Replace the damaged
items.
Metal Exposed
Is the insulation of the probe or connection
cable to be used damaged, or is bare metal
exposed?
When using probes and connection cables
When using a clamp
Is the clamp cracked or damaged?
Yes
1
No Metal Exposed
2
Go to
No
2
Go to
Peripheral Device Inspection
If damage is evident, request re-
pairs.
Yes
Is damage to the instrument or input mod-
ules evident?
Instrument and Input Module Inspection
When turning power on
Do the fans rotate and the Hioki logo appear
on the screen?
No
2
The power cord may be damaged,
or the instrument may be dam-
aged internally. Request repairs.
Does the Waveform screen appear?
No
Yes
The instrument may be damaged
internally. Request repairs.
Nothing appears,
or the display is
abnormal
Yes
Inspection complete
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......