12.1 Overview and Features
243
11
Chapte
r 12 FF
T Function
12
FFT analysis can only be used with the FFT function.
The FFT (Fast-Fourier Transform) functions provide frequency analysis of input signal data.
Use these functions for frequency analysis of rotating objects, vibrations, sounds and etc.
For details, refer to "Appendix 4 FFT Definitions" (p.A13).
Analysis can be performed on data as it is being measured, on pre-existing analog waveform data previ-
ously acquired with the Memory function, and on data output from waveform calculations.
When using Model 8968 High Resolution Unit equipped with an anti-aliasing filter, the cut-off frequency
can be automatically set by linking with the frequency range setting.
Major Features
•FFT analysis frequency range: 133 mHz to 8 MHz
•FFT Analysis Modes (16 types)
* Not available when using external sampling.
For phase spectra, only the required phase information is highlighted and displayed.
See:
"12.3.8 Emphasizing Analysis Results (phase spectra only)" (p.255)
so, when performing FFT analysis with the instrument connected to a sound level or vibration meter, scal-
ing by dB can be set from the Channel Settings screen if you want to read values directly in calibrated
units of measurement.
See:
Refer to the "17.2.3 FFT Function" (p.343) for FFT function specifications.
FFT Function
Chapter 12
12.1 Overview and Features
• Storage Waveform
• Cross-power Spectrum
• Histogram
• Impulse Response
• Linear Spectrum
• Coherence Function
• RMS Spectrum
• Phase Spectrum
• Power Spectrum
• Auto-correlation Function
• Power Spectrum Density*
• Cross-correlation Function
• LPC analysis (Power Spectrum
Density)
*
• 1/1 Octave Analysis*
• Transfer Function
• 1/3 Octave Analysis*
To suppress the effects of aliasing distortion
• We recommend using Model 8968 High Resolution Unit that are equipped with anti-aliasing filter-
ing to suppress the effects of aliasing distortion when sampling.
• FFT analysis cannot be performed for MR8990 Digital Voltmeter Unit.
See:
Aliasing Distortion and Anti-Aliasing Filters
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......