18.7 Disposing of the Instrument (Lithium Battery Removal)
377
C
hap
ter 18
M
aintena
nce and
Service
14
18
The instrument contains a lithium battery for memory backup. Remove this battery before disposing of the
instrument.
Procedure ____________________________________________________
Required tools:
Phillips screwdriver, wire cutter: One each
Box wrench or long-nosed pliers
1.
Verify that the power is off, and remove the connection cables and power cord.
2.
Remove the screws and each parts as indicated in the next page diagram.
3.
Pull the lithium battery up from the circuit board, and cut the two leads with a
wire cutter.
4.
Remove the battery from the board.
18.7 Disposing of the Instrument (Lithium Battery
Removal)
• To avoid electric shock, turn off the power switch and disconnect the power
cord and connection cords before removing the lithium battery.
• Keep batteries away from children to prevent accidental swallowing.
• Battery may explode if mistreated. Do not short-circuit, recharge, disassemble
or dispose of in fire.
When disposing of this instrument, remove the lithium battery and dispose of
battery and instrument in accordance with local regulations.
CALIFORNIA, USA ONLY
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......