7.9 Making Detailed Settings for Input Modules
177
6
Chapte
r 7 Utility Functions
7
Offset
For DC output: Sets DC voltage.
For sine wave output: Sets the offset voltage.
The output voltage with guaranteed accuracy is the sum of the amplitude and the
offset, between -10 V and +10 V. If the sum of the amplitude and the offset is set
outside the guaranteed accuracy range, parts of the waveform will be clamped to
the upper limit, approxi16 V and the lower limit, approximately -11 V.
Phase
Sets the phase.
Duty
Sets the duty when selecting pulse wave.
When Off
Sets the output terminal status when the output is off.
Select
Output
Turns waveform output On/Off.
Select
Control
Sets the waveform output.
Select
-10 V to +15 V
-360
°
to 360
°
0.1% to 99.9%
Selections
Description
Open
Opens the output terminals, separating them from internal circuits.
Short
Short-circuits the output terminals, separating them from internal
circuits.
Selections
Description
On
Outputs waveform.
Off
Does not output the waveform.
Selections
Description
RUN
Starts output. (Output indicator: Red)
PAUSE
Pauses output. While output is paused, the output at the time
[PAUSE]
was pressed will be output. (Output indicator: Red)
STOP
Stops output. (Output indicator: Off)
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......