5.2 Setting Basic Settings of Measurement
158
The value of the test signal level may change according to the sample which is being tested.
This instrument is possible to vary the level of the test signal applied to the object under test over a wide
range using the following three methods. Selecting constant voltage or constant current mode will result in
increased measurement times due to use of software feedback control.
5.2.6 Setting the Measurement Signal Level
Do not switch between V, CV and CC while the test sample is still connected to the
measurement terminals because doing so may damage the test sample.
Open circuit voltage (V) mode
The value of the open circuit voltage is set.
Constant voltage (CV) mode
The value of the voltage between the terminals of the object under
test is set.
Constant current (CC) mode
The value of the current flowing through the object under test is set.
Procedure
ANALYZER Measurement Screen
ANALYZER Basic Settings
1
Press .
ANALYZER Basic Settings