12.2 Timing Chart
321
12
Chapte
r 1
2 External Cont
rol
Touching
on the screen asking whether to input the trigger signal from EXT I/O in ANALYZER
mode will result in the following output.
Furthermore, if the panel number is selected with the panel load signal when a trigger signal is input from the
EXT I/O, the measurement condition of that panel number is loaded and then measurement is performed.
The following shows examples of the measurement timing.
(In the timing examples, the valid edge of the TRIG signal is set to falling (ON).)
12.2.2 ANALYZER Mode (IM3533-01 only)
*1: Reset when signal changes to EOM (HIGH): HIGH
Do not reset when signal changes to EOM (HIGH): Last judgment result remains
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
Judgment result
(Error output)
Last judgment
result
Signal line
Description
INDEX
The transition to HIGH is performed when measurement of the first sweep point starts after trig-
ger signal input and the transition to LOW is performed when the analog measurement of the last
sweep point ends.
(The HIGH level is maintained during sweep measurement.)
EOM
The transition to HIGH is performed when measurement of the first sweep point starts after trig-
ger signal input. Measurement of the last sweep point ends and the transition to LOW is per-
formed after judgment result output.
(The HIGH level is maintained during sweep measurement.)
• If the trigger setting is set to STEP, INDEX and EOM transition to LOW every time the
measurement for each point ends, and then transition to HIGH when there is trigger
input.
• For other timing chart times, refer to "12.2.1 LCR Mode" (p. 318).
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......