This chapter provides example measurement scenarios for LCR mode, ANALYZER mode (IM3533-01 only),
and TRANSFORMER mode.
Necessary items: 9263 SMD test fixture, Laminated ceramic capacity you want to measure
Measurement
Example
Chapter 3
3.1 When LCR Mode
Measuring a Laminated Ceramic Capacitor
1
Connect the 9263 SMD test fixture to the measurement terminals.
For the connection procedure, refer to the
instruction manual supplied with the fixture.
2
Set the first parameter to Cs and the third parameter to D. (p. 46)
Trigger synchronous output function
setting: OFF (p. 77)
3
Set the measurement conditions.
Touch
on the Measurement screen, select
the item you want to set, and set it as follows.
LCR Basic Settings
Measurement frequency: 1.0000 kHz
(p. 50)
Measurement signal mode:
Open circuit voltage (V) mode (p. 52)
Measurement signal level: 1.000 V (p. 52)
Measurement range: AUTO (p. 62)
Measurement speed: MED (p. 73)
Voltage and current limit: (p. 56)
Average: OFF (p. 74)
Trigger delay: 0.0000 s (p. 76)
Trigger: INT(p. 60)
DC bias: OFF (p. 58)
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......