14.1 General Specifications
338
Output impedance
(Hc terminal, when 1 kHz)
Normal mode: 100
±
10
Low impedance accuracy mode: 25
±
5
Measurement signal level
(1) Open circuit terminal voltage (V) mode and constant voltage (CV) mode
• Level range
Normal mode: 5 mV to 5 V, maximum 50 mA
Low Z high accuracy mode: 5 mV to 2.5 V, maximum 100 mA
• Setting resolution 1 mV steps
• Setting accuracy
±
10%
of setting
±
10 mV
(2) Constant current (CC) mode
• Level range
Normal mode: 10
A to 50 mA, maximum 5 V
Low Z high accuracy mode: 10
A to 100 mA, maximum 2.5 V
• Setting resolution 10
A steps
• Level accuracy
±
10%
of setting
±
10
A
Measurement range
The measurement range is determined according to impedance Z.
The values of the other measurement items can be calculated.
Ranges: 100 m
, 1
, 10
, 100
, 1 k
, 10 k
, 100 k
, 1 M
, 10 M
, 100 M
(10 ranges)
• The guaranteed accuracy range differs depending on the measurement conditions
• Out of guaranteed accuracy is displayed when out of the ranging range
OVERFLOW or UNDERFLOW is displayed when out of the A/D input range
Low Z High Accuracy Mode • Improves measurement accuracy in the 100 m
and 1
ranges.
The measurement current is increased (maximum 100 mA, maximum applied voltage 2.5
V) and the measurement accuracy is improved by setting the output resistance to 25
.
• Low Z high accuracy mode is only available for the 100 m
and 1
ranges.
1.Basic Specifications
Measurement range
Guaranteed Accuracy Range
AUTO Ranging Range
100 M
8 M
to 200 M
8 M
to
10 M
800 k
to 100 M
800 k
to 10 M
1 M
80 k
to 10 M
80 k
to 1 M
100 k
8 k
to 1 M
8 k
to 100k
10 k
800
to 100 k
800
to 10 k
1 k
80
to 10 k
80
to 1 k
100
8
to 100
8
to 100
10
800 m
to 10
800 m
to 10
1
80 m
to 1
80 m
to 1
100 m
10 m
to 100 m
0
to 100 m
Setting range of low Z high accuracy mode
No
Measurement
range
to 1 kHz
to 10 kHz
to 100 kHz
to 200 kHz
1
100 M
2
10 M
3
1 M
4
100 k
5
10 k
6
1 k
7
100
8
10
9
1
10
100 m
Low Z high accuracy mode/ normal mode
None
Normal mode only (low Z high accuracy mode setting disabled)
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......