4.5 Setting Application Settings
133
4
Chapte
r 4
LCR Function
•
Selecting
or
as the contact check timing causes the trigger synchro-
nous output function to be automatically turned on.
See
•
When setting the contact check function, the INDEX time and EOM time will be delayed
depending on the timing. (p. 352)
•
The contact check function cannot be used during temperature measurement. However, if
a contact error occurs at the
timing, the temperature measurement will be shown
as
DISP OUT
since the measurement will not have been performed at that point.
•
The allowable contact resistance value may vary depending on the sample being mea-
sured.
•
The measurement value will not be saved when all three of the following conditions apply:
• When the memory function has been enabled
• When the contact check timing is set to
• When a contact check error has been displayed (see the error display) (p. 360)
Set the contact check threshold with
and
.
4
Settable range : 1 to 5
Contact Check Settings
Threshold
1
2
3
4
5
Permissible
contact
resistance [
]
Approx.
1000
Approx.
500
Approx.
100
Approx.
50
Approx.
10
5
Press
to close the setting screen.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......