8.1 Setting Open Circuit Compensation
222
A window such as the following will be displayed if the instrument was unable to acquire normal compensation
values. If this occurs, the acquired compensation values can be enabled by touching
.
However, those values are not guaranteed.
The open circuit compensation process is quite sensitive to noise - both noise originating externally and
induced noise.Therefore, if open circuit compensation has been interrupted with a fault, you should check
the following points before starting the compensation process again:
• Check that the test cables are properly connected.
• Check that nothing is connected to the test cables. (Open circuit compensation cannot be performed
while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near
them.
• Execute the guarding process.
"Appendix2 Measurement of High Impedance Components"(p. A3)
When Normal Compensation Values were Not Acquired
Screen when Normal Compensation Values were Not Acquired
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......