14.2 Measurement Range and Accuracy
349
14
Chapter 14
S
pecifica
tions
Pt sensor
Sheath-type temperature probe
HIOKI 9478
Measurement conditions
At least 60 min. after power supply activation
Temperature and humidity: 23±5°C, 80% rh or less
Accuracy
Conversion table from C and L to |Z|
Frequency (Hz)
Temperature measurement
Guaranteed accuracy range
-10.0°C to 99.9°C
Accuracy
±0.5%rdg±1°C
*1
Sampling time
Approx. 640 ms
*1 Indicates the accuracy of the instrument used in combination with the 9478 Sheath Type Temperature Probe.
For ambient temperatures of 0°C to 18°C and 28°C to 40°C, add 0.02°C/°C to the measurement accuracy.