8.3 Compensating Values to Match Reference Values (Load Compensation)
232
Compensate measurement values to match the element that will be the reference.
With load compensation it is possible to calculate the compensation coefficient by measuring a reference
sample with known data and perform the compensation for the test data obtained from the target sample.
The compensation coefficient can be acquired using up to five compensation conditions. The reference value
of each compensation condition can be set independently.
The following seven compensation conditions should be set for each measurement frequency:
The compensation coefficient is computed from the reference values of Z and
obtained from the set values
and the actual data acquired from the reference sample at each of the compensation frequencies.
Compensation coefficient of Z =
Compensation value of
= (Reference value of
) - (Actual data of
)
The measured values of Z and
are first compensated using the following equations, and then individual
parameters from the compensated Z and
values are employed.
Z = (Z before compensation) × (Compensation coefficient of Z)
= (
before compensation) + (Compensation value of
)
8.3 Compensating Values to Match
Reference Values (Load Compensation)
Set the Z/ Cs/ Cp/ Ls/ Lp/ Rs reference
value selected for the parameter type
(p. 242).
Define the measurement frequency
used to measure and compensate the
reference sample. (p. 237)
Set the range to compensate. (p. 238)
Set the type and value of the
measure-
ment
signal mode to compensate.
DC Bias
Enable or disable DC bias and set the
value. (p. 240)
Set the parameter to use as the refer-
ence value. (p. 241)
Reference
Value 2
Deletes the compen-
sation conditions
(p. 243).
Reference
Value 1
Parameter Type
Compensation
Signal Level
Compensation
Range
Compensation
Frequency
Set the
/ D/ Rs/ Rp/ Q/ X reference
value selected for the parameter type
(p. 242).
Loads the current measurement
conditions as compensation
conditions (p. 243).
(Actual data of Z)
(Reference value of Z)
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......