4.5 Setting Application Settings
137
4
Chapte
r 4
LCR Function
The higher the measurement frequency, the shorter the time that INDEX and EOM are high (off). When the
high (off) time is too short due to characteristics of the input circuit, the instrument can be configured to
maintain the low (on) state for a preset time once EOM changes to low (on) before reverting the signal to high
(off) after the completion of measurement. The INDEX output method can be changed in the same manner.
See
"Chapter 12 External Control" (p. 309)
4.5.7
Setting the EOM Output Method
Procedure
LCR Measurement Screen
LCR Application Settings
1
Press
.
2
LCR Application Settings
Setting the output method.
Use
or
to set the EOM output time
for the PULSE setting.
For HOLD and PULSE timing charts, see
"Chapter 12 External Control" (p. 309).
3
Setting the EOM output method
• Settable range: 0.0001 s to 0.9999 s
• If you make a mistake during input:
press
to cancel the input and start
again.
• The output time cannot be set unless the
output method has been set to [PULSE].
4
Press
to close the setting screen.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......