Before requesting instrument repair or inspection, please read "Before returning for repair" (p. 355) and
Section "Error display" (p. 360).
Certain parts require replacement periodically and at the end of their useful life: (Useful life depends on
the operating environment and frequency of use. Operation cannot be guaranteed beyond the following
periods)
Maintenance
and Service
Chapter 15
15.1 Inspection, Repair and Cleaning
Inspection and Repair
Do not attempt to modify, disassemble or repair the instrument; as fire, electric shock
and injury could result.
•
If damage is suspected, check the "Before returning for repair" (p. 355) section before
contacting your dealer or Hioki representative.
•
If the fuse blows, do not attempt to replace the fuse or repair the instrument: contact your
authorized Hioki distributor or reseller.
•
If damage is suspected, check the"Before returning for repair" (p. 355) section before
contacting your dealer or Hioki representative.
However, in the following cases, immediately stop using the instrument, unplug the
power cord and contact your authorized Hioki distributor or reseller.
• When the nature of the damage is clearly evident
• When measurement is impossible
• After long-term storage in adverse conditions such as high temperature or humidity
• After being subject to severe shock during transport
• After severe exposure to water, oil, or dust (internal insulation can be degraded by oil
or water, causing increase hazard of electric shock or fire)
Replaceable Parts
Part
RemarksLife
Remarks
Electrolytic
Capacitors
Approx. 10 years
The useful life of electrolytic capacitors depends on the operating environ-
ment.Periodic replacement is necessary.
Lithium battery
Approx. 10 years
The instrument incorporates a lithium battery for backup. The life of the back-
up battery is approximately 10 years. If the date and time greatly differ from
the actual date and time when the power is turned on or a backup error ap-
pears at startup, it is time to replace the battery.
Contact your authorized Hioki distributor or reseller.
LCD backlight
(to half brightness)
Approx. 50,000
hours
Periodic replacement is necessary.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......