15.3 Error display
360
When any of the following indications appear on the screen, check the corresponding reference page.
15.3 Error display
Error display
Description
Remedy and Reference
The life of the RAM backup battery has
ended.
The instrument needs to be repaired.
Contact your authorized Hioki distributor
or reseller.
This is displayed when a measurement
value is outside of the guaranteed accu-
racy range.
Increase the measurement signal level
or change the measurement range to
one that matches the impedance of the
element to be measured (p. 52), (p. 62).
This is displayed when load compensa-
tion is enabled and the load compensa-
tion frequency does not match the
current measurement frequency.
Match the current measurement fre-
quency to the compensation frequency
(p. 232).
When LCR mode:
This is displayed when constant voltage
measurement and constant current
measurement cannot be performed.
Reduce the constant voltage level or
constant current level (p. 54).
When ANALYZER mode:
This is displayed when a signal level
that is lower than the set value is applied
to the test sample as a result of the volt-
age/current limit value setting.
Set the limit value again or change the
measurement signal level so that the
limit value is not exceeded (p. 56).
This is displayed when load compensa-
tion is enabled and a load compensation
condition other than the frequency does
not match the current measurement
condition.
Match the current measurement condi-
tion to the load compensation condition
(p. 232).
This is displayed when the set number
of measurement results have been
stored in the instrumentÅfs memory.
Load measurement values stored in the
instrument’s memory with the memory
function or clear the memory. (p. 138)
• This is displayed when a measure-
ment value is outside of the screen
display range.
• This is displayed when no tempera-
ture probe is connected during tem-
perature measurement.
• Change the measurement range to
one that matches the impedance of
the element to be measured (p. 62).
• Check the measurement probe con-
nection. (p. 33)
This is displayed when measurement
does not end because of an internal cir-
cuit error.
The instrument needs to be repaired.
Contact your authorized Hioki distributor
or reseller.
This is displayed when a measurement
value is at or above the upper limit value
of the auto ranging range.
Change the measurement range to a
high-impedance range (p. 62).
This is displayed when a measurement
value is at or below the lower limit value
of the auto ranging range.
Change the measurement range to a
low-impedance range (p. 62).
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......