6.1 About TRANSFORMER function
195
6
Chapte
r 6
TRANSFORMER Functio
n
Connect the two sides in phase and in series and touch
. Then connect them in reverse phase and
in series and touch
to display the mutual inductance.
Mutual inductance
Procedure
1
Connect the two sides in phase and in series as
shown below and measure with
.
TRANSFORMER Measurement Screen
Primary
side
Secondary
side
In phase, in series connection
In-phase, in-series
measurement value
Primary
side
Secondary
side
2
Connect the two sides in reverse phase and in series
as shown below and measure with
.
TRANSFORMER Measurement Screen
Reverse phase, in series connection
Reverse-phase, in-series
measurement value
Calculated value