Appendix4 Countermeasures Against Incorporation of External Noise
A5
The unit is designed to be resistant to errors caused by interference from the test cables or the power supply
line. However, if the level of the interference is particularly large, this can cause measurement errors or faulty
operation.
Refer to the examples given below for examples of countermeasures which can be taken against interference
which has caused faulty operation etc.
You can use the following countermeasures to reduce the effect of noise being incorporated from the power
line.
The unit is structured so that the ground wire of the power cable can be used as protective grounding for the
unit. Protective grounding plays an important role in not only the prevention of electrical accidents but also the
use of an internal filter to eliminate the incorporation of noise from the power line. Use the supplied power
cord.
Connect a commercial plug-in noise filter to the power outlet and then connect the unit to the output of the
noise filter in order to suppress the incorporation of noise from the power line.
Plug-in noise filters are commercially available from various specialist manufacturers.
Appendix4 Countermeasures Against Incorpora-
tion of External Noise
Appendix4.1 Countermeasures Against Incorporation
of Noise from the Power Line
Grounding Using a Protective Ground Wire
Attaching a Noise Filter to the Power Line
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......