12.1 External Input/Output Connector and Signals
312
32
OUT
BIN6, PARA3-LO
BIN judgment results.
Outputs LO comparator judgment results for the No. 3 parameter.
Neg
Level
33
OUT
BIN8
BIN judgment results
Neg
Level
34
OUT
BIN10
BIN judgment results
Neg
Level
35
-
(Unused)
-
-
-
36
-
(Unused)
-
-
-
37
-
(Unused)
-
-
-
Pin
I/O
Signal name
Function
Logic
ANALYZER mode (IM3533-01 only)
Pin
I/O
Signal name
Function
Logic
1
IN
TRIG
External trigger (p. 316)
Pos/
Neg
Edge
2
-
(Unused)
-
-
-
3
-
(Unused)
-
-
-
4
IN
LD1
Select panel number (p. 316)
Neg
Level
5
IN
LD3
Select panel number (p. 316)
Neg
Level
6
IN
LD5
Select panel number (p. 316)
Neg
Level
7
-
(Unused)
-
-
-
8
-
ISO_5V
Isolated 5 V power output
-
-
9
-
ISO_COM
Isolated common signal ground
-
-
10
OUT
ERR
Outputs when a sampling error, overcurrent error, contact error,
HIGH-Z reject error, or constant voltage/constant current error oc-
curs.
Neg
Level
11
-
(Unused)
-
-
-
12
-
(Unused)
-
-
-
13
-
(Unused)
-
-
-
14
-
(Unused)
-
-
-
15
-
(Unused)
-
-
-
16
-
(Unused)
-
-
-
17
-
(Unused)
-
-
-
18
-
(Unused)
-
-
-
19
-
(Unused)
-
-
-
20
-
(Unused)
-
-
-
21
-
(Unused)
-
-
-
22
IN
LD0
Select panel number (p. 316)
Neg
Level
23
IN
LD2
Select panel number (p. 316)
Neg
Level
24
IN
LD4
Select panel number (p. 316)
Neg
Level
25
IN
LD6
Select panel number (p. 316)
Neg
Level
26
IN
LD_VALID
Neg
Level
27
-
ISO_COM
Isolated common signal ground
-
-
28
OUT
EOM
Measurement complete signal: When this signal is output, the
comparator judgment results have been finalized.
Neg
Edge
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......